【Industrial Scientific 常见问题和回答】
我应当使用何种流速来校正我的仪器?
应当多长时间校正仪器?
如果使用SP402或者SP202采样泵,我可以吸取多远的密闭空间内样品?
使用采样泵可以提高仪器的测量速度吗?
仪器的电池类型是怎样的?
一起充满电可以用多长时间?
TWA 和STEL代表什么?
Industrial Scientific有固定式的检测仪器吗?
传感器的寿命如何?
Who are your distributors and where are they located?
What are the filters that come with the sampling pump used for?
How do I dispose of my empty calibration bottles (cylinders)?
Why do you recommend Pentane calibration gas over the use of Methane?
What accessories does Industrial Scientific have to make my instruments perform better in a variety of applications?
 
【国际认证】
→What flow rate should I use for calibrating my gas detection instrument?
·For most gases a flow of .5 lpm (1 SCFH) is sufficient. However, for chlorine, ammonia, nitrogen dioxide, sulfur dioxide and hydrogen cyanide a flow rate of 1.0 lpm (2 SCFH) may be required. Please refer to your instrument instruction manual for recommended flow rates for a specific instrument.
→How often should I calibrate?
·Gas detection instruments are potential life-saving devices. Recognizing this fact, Industrial Scientific Corporation recommends that a functional ("bump") test be performed on every instrument prior to each day's use. A functional test is defined as a brief exposure of the monitor to a known concentration of gas(es) for the purpose of verifying sensor and alarm operation and is not intended to be a measure of the accuracy of the instrument.
·Industrial Scientific further recommends that a full instrument calibration be performed using a certified concentration(s) of calibration gas(es) monthly to ensure maximum accuracy.
·If an instrument fails to operate properly following any functional "bump" test, a full instrument calibration should be performed prior to use.
→How far can I draw a gas sample from a confined space with the SP402 or SP202 sampling pumps?
·You can draw a sample from a distance of 100ft. (30.5 m) using either 1/8" Tygon or 3/16" Tygon lined w/Teflon tubing.
 
→What type of battery powers the instruments?
·The typical power source for our gas detection instruments is rechargeable Ni-Cad. Depending on your instrument choice, there other options as well. For 200 Series instruments several models use disposable alkaline batteries, and our 300 and 400 Series instruments can utilize a disposable Lithium battery pack. The iTX and VX500 instruments offer both Alkaline and rechargeable Lithium-ion batteries.
 
→How long do the instruments run on full charge?
·Typical run time of our instruments is 10 hours. For the 300 & 400 Series instruments using a Lithium battery pack the run time jumps up to 14 hours.
 
→What are TWA and STEL?
·TWA stands for Time Weight Average. This in reference to dosages of toxic gas you may encounter in the work place. It is based on an 8 hour day/ 40 hour work week. TWA ia a term established by the American Conference of Governmental Industrial Hygienists (ACGIH).
·STEL stands for Short Term Exposure Limit. This is the average amount of gas you can be exposed to in a 15 minute period with no long term health effects. This may occur 4 times a day. STEL is a term established by the ACGIH.
→Does Industrial Scientific have fixed point gas detection systems?
·Yes. Industrial Scientific has different fixed-point monitoring systems for many gases and many applications. Please visit our fixed systems product section of the website Fixed Monitoring Systems.
 
→How long do sensors last?
·Sensors used by Industrial Scientific are guaranteed for 1 year with the exception of the ammonia sensor (it is guaranteed 6 months). Typical life is much greater than that. A combustible sensor usually lasts in excess of 5 years. Typical life of CO and H2S sensors are 2-3 years. The MDU420 patented infrared sensor is warranted for a period of 5 years.
→Who are your distributors and where are they located?
·We have a distribution network throughout the world. Our representatives are factory trained in the operation and maintenance of our products and handle a complimentary line of safety products. To locate the nearest Industrial Scientific distributor, click on the Distributor Locator link Distributor Locator or E-mail info@indsci.com.
→What are the filters that come with the sampling pump used for?
·There are two filters delivered with Industrial Scientific sampling pumps. On top of the pumps is a 10 micron dust filter. This is required for the protection of the internal workings of the motorized pump. In addition a dust filter/water stop is supplied. This filter is a .45 micron filter and it protects the unit from smaller dust particles. It also stops any liquid from entering the sampling tube, which in turn protects the sampling pump from any damage if such liquids as water, gasoline or paint thinner are present in the sampling area.
→How do I dispose of my empty calibration bottles (cylinders)?
·The gas calibration cylinders supplied by Industrial Scientific are disposable. The empty cylinders may be thrown away as regular trash if vented of all remaining pressure. Remove the regulator and dispose of the cylinder. Some companies are required to dispose of these cylinders as hazardous waste, either by company policies or state regulations. If your company falls into this category, simply return the empty cylinders to Industrial Scientific.
→Why do you recommend Pentane calibration gas over the use of Methane?
·Methane Calibration has been the standard for years. However, the typical confined space may contain gases other than methane. Using pentane as a calibration standard will adjust the combustible gas sensor in the Lower Explosive Level (LEL) mode to detect a larger group of hydrocarbons. The instrument will keep the gas reading in a safe level. For a correlation chart contact the customer service group.
→What accessories does Industrial Scientific have to make my instruments perform better in a variety of applications?
·Industrial Scientific has a large variety of accessories to use with our complete instrument line. Look in the accessory section of the Industrial Scientific Homepage, or contact our customer service group.
 
 
 